Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, relative humidity at 0 % to 75 %. Accuracy specifications take the form of: ±(% of Reading + Number of Least Significant Digits).
Function | Range | Resolution | Accuracy | |
AC volts (40 Hz to 500Hz)1 |
4.000 V 40.00 V 400.0 V 1000 V |
0.001 V 0.01 V 0.1 V 1 V |
1.0 % + 3 | |
DC volts |
4.000 V 40.00 V 400.0 V 1000 V |
0.001 V 0.01 V 0.1 V 1 V |
0.5 % + 3 | |
AC millivolts | 400.0 mV | 0.1 mV | 3.0 % + 3 | |
DC millivolts | 400.0 mV | 0.1 mV | 1.0 % + 10 | |
Diode test2 | 2.000 V | 0.001 V | 10% | |
Resistance (Ohms) |
400.0 Ω 4.000 kΩ 40.00 kΩ 400.0 kΩ 4.000 MΩ 40.00 MΩ |
0.1 Ω 0.001 kΩ 0.01 kΩ 0.1 kΩ 0.001 MΩ 0.01 MΩ |
0.5 % + 3 0.5 % + 2 0.5 % + 2 0.5 % + 2 0.5 % + 2 1.5% + 3 |
|
Capacitance3 |
40.00 nF 400.0 nF 4.000 μF 40.00 μF 400.0 μF 1000 μF |
0.01 nF 0.1 nF 0.001 μF 0.01 μF 0.1 μF 1 μF |
2 % + 5 2 % + 5 5 % + 5 5 % + 5 5 % + 5 5 % + 5 |
|
Frequency1 Hz (10 Hz – 100 kHz) |
50.00 Hz 500.0 Hz 5.000 kHz 50.00 kHz 100.0 kHz |
0.01 Hz 0.1 Hz 0.001 kHz 0.01 kHz 0.1 kHz |
0.1 % + 3 | |
Duty Cycle1 | 1 % to 99 % | 0.1 % | 1 % typical 4 | |
AC current μA (40 Hz to 400 Hz) |
400.0 μA 4000 μA |
0.1 μA 1 μA |
1.5 % + 3 | |
AC current mA (40 Hz to 400 Hz) |
40.00 mA 400.0 mA |
0.01 mA 0.1 mA |
1.5 % + 3 | |
AC current A (40 Hz to 400 Hz) |
4.000 A 10.00 A |
0.001 A 0.01 A |
1.5 % + 3 | |
DC current μA |
400.0 μA 4000 μA |
0.1 μA 1 μA |
1.5 % + 3 | |
DC current mA |
40.00 mA 400.0 mA |
0.01 mA 0.1 mA |
1.5 % + 3 | |
DC current A |
4.000 A 10.00 A |
0.001 A 0.01 A |
1.5 % + 3 | |
Temperature |
50 °C- 400 °C 0 °C- 50 °C -55 °C- 0 °C |
0.1C |
2 % ± 1 °C ± 2 °C 9 % ± 2 °C |
|
Backlight | - | - | Yes | |
1All ac, Hz, and duty cycle are specified from 1 % to 100 % of range. Inputs below 1 % of range are not specified. 2Typically, open circuit test voltage is 2.0 V and short circuit current is <0.6 mA. 3Specifications do not include errors due to test lead capacitance and capacitance floor (may be up to 1.5 nF in the 40 nF range). 4Typical means when the frequency is at 50 Hz or 60 Hz and the duty cycle is between 10 % and 90 %. |
Function | Overload Protection | Input Impedance (Nominal) | Common Mode Rejection Ratio | Normal Mode Rejection Ratio |
AC volts | 1000 V 1 | >10 MΩ <100 pF |
>60 dB at dc, 50 Hz or 60 Hz |
- |
AC millivolts | 400 mV | >1MΩ, <100 pF | >80 dB at 50 Hz or 60 Hz | - |
DC volts | 1000 V 1 | >10 MΩ <100 pF |
>100 dB at dc, 50 Hz or 60 Hz |
>60 dB at 50 Hz or 60 Hz |
DC millivolts | 400 mV | >1MΩ, <100 pF | >80 dB at 50 Hz or 60 Hz | - |
1 106 V Hz Max |
General Specifications | |
Maximum voltage between any terminal and earth ground | 1000 V |
Display (LCD) | 4000 counts, updates 3/sec |
Battery type | 2 AA, NEDA 15A, IEC LR6 |
Battery life | 500 hours minimum (50 hours in LED Test mode without load. The hours with load depends on the type of LED under test.) |
Temperature | |
Operating | 0 °C to 40 °C |
Storage | -30 °C to 60 °C |
Relative Humidity | |
Operating humidity |
Non-condensing (<10°C) ≤90% RH at 10 °C to 30 °C; ≤75% RH at 30 °C to 40 °C |
40 MΩ range ≤80% RH at 10 °C to 30 °C; ≤70% RH at 30 °C to 40 °C |
|
Altitude | |
Operating | 2000 m |
Storage | 12,000 m |
Temperature coefficient | 0.1 X (specified accuracy) /°C (<18 °C or >28 °C) |
Fuse protection for current inputs |
440 mA, 1000 V Fast Fuse, Fluke specified part only. 11A, 1000V Fast Fuse, Fluke specified part only. |
Size (HxWxL) | 183 x 91 x 49.5 mm |
Weight | 455 g |
IP rating | IP 40 |
Safety | IEC 61010-1, IEC61010-2-030 CAT III 600 V, CAT II 1000 V, Pollution Degree 2 |
Electromagnetic environment | IEC 61326-1: Portable |
Electromagnetic compatibility | Applies to use in Korea only |
Class A equipment (industrial broadcasting & communication equipment)1 1 This product meets requirements for industrial (Class A) electromagnetic wave equipment and seller or user should take notice of it. This equipment is intended for use in business environments and is not to be used in homes. |
Công ty Cổ Phần Thương mai Dịch vụ và Xuất Nhập Khẩu Bình Minh
Địa chỉ: Số 9,Ngõ 461 Nguyễn Văn Linh, Q. Long Biên, TP.Hà nội
Tel: 024-38751616 Hotline: 0904499667
Email: [email protected]
============================
Chi nhánh Công ty Cổ Phần Thương mai Dịch vụ và Xuất Nhập Khẩu Bình Minh Tại TPHCM
Địa chỉ: Số 1331/15/144 Đường Lê Đức Thọ,Phường 14, Q. Gò Vấp, TP.Hồ Chí Minh
Tel: 024-38751616 và 024-36790555 Hotline: 0988764055
Email: [email protected]
AMPE KÌM |
ĐỒNG HỒ ĐO ĐIỆN Đồng hồ đo điện trở cách Kyoritsu Đồng hồ đo điện trở cách Hioki Đồng hồ đo điện trở đất Kyoritsu
|
MÁY THỬ ĐIỆN ÁP |
MÁY ĐO NHIÊT ĐỘ
|
MÁY ĐO TỐC ĐỘ |