PROVA 21 Click vào Tải Thông Số Kỹ Thuât Chi Tiên
	
	PROVA 21 Power Harmonics and Leakage Tester
Electrical Specifications: (23℃±5℃)
AC Watt
(50 or 60 Hz, PF 0.6~1, CT=1, Voltage is greater than AC 4V, Current is greater than AC 1mA for mA range, and Current is greater than AC 0.04A for A range. Specification applies to continuous waveforms)
| 
				 Range (0 - 30A)  | 
			
				 Resolution  | 
			
				 Accuracy of Readings  | 
		||
| 
				 0.050 – 9.999 W  | 
			
				 0.001W  | 
			
				 ±2% ± 0.025W  | 
		||
| 
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
		|
| 
				 10.00 – 99.99 W  | 
			
				 0.01W  | 
			
				 ±2% ± 0.25W  | 
		||
| 
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
		|
| 
				 100.0 - 999.9 W  | 
			
				 0.1W  | 
			
				 ±2% ± 2.5W  | 
		||
| 
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 1.000  | 
			
				 – 9.999 KW  | 
			
				 0.001 KW  | 
			
				 ±2% ± 0.025KW  | 
		|
| 
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 10.00  | 
			
				 – 99.99 KW  | 
			
				 0.01 KW  | 
			
				 ±2% ± 0.25KW  | 
		|
| 
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 100.0  | 
			
				 – 999.9 KW  | 
			
				 0.1 KW  | 
			
				 ±2% ± 2.5KW  | 
		|
| 
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 1000  | 
			
				 – 9999 KW  | 
			
				 1 KW  | 
			
				 ±2% ± 25KW  | 
		|
| 
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
		|
| 
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
		|
| 
				 Range (30 – 50A)  | 
			
				 
  | 
			
				 Resolution  | 
			
				 Accuracy  | 
		|
| 
				 0.050 – 9.999 W  | 
			
				 
  | 
			
				 0.001W  | 
			
				 ±2% of VA ± 5dgts  | 
		|
| 
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
		|
| 
				 10.00 – 99.99 W  | 
			
				 
  | 
			
				 0.01W  | 
			
				 ±2% of VA ± 5dgts  | 
		|
| 
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
		|
| 
				 100.0 - 999.9 W  | 
			
				 
  | 
			
				 0.1W  | 
			
				 ±2% of VA ± 5dgts  | 
		|
| 
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 1.000  | 
			
				 – 9.999 KW  | 
			
				 
  | 
			
				 0.001 KW  | 
			
				 ±2% of VA ± 5dgts  | 
		
| 
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 10.00  | 
			
				 – 99.99 KW  | 
			
				 
  | 
			
				 0.01 KW  | 
			
				 ±2% of VA± 5dgts  | 
		
| 
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 100.0  | 
			
				 – 999.9 KW  | 
			
				 
  | 
			
				 0.1 KW  | 
			
				 ±2%of VA ± 5dgts  | 
		
| 
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 1000  | 
			
				 – 9999 KW  | 
			
				 
  | 
			
				 1 KW  | 
			
				 ±2% of VA ± 5dgts  | 
		
AC Voltage
(Auto Range, True RMS, Crest Factor<4, Input Impedance 10 MΩ, Max. peak value 860VDC, and Overload Protection AC 800V)
| 
				 Range  | 
			
				 Resolution  | 
			
				 Accuracy of Readings (50/60Hz)  | 
		
| 
				 5V – 250 V  | 
			
				 0.1 V  | 
			
				 ±0.5% ± 5dgts  | 
		
| 
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 250 V – 600 V  | 
			
				 0.1 V  | 
			
				 ±0.5% ± 5dgts  | 
		
| 
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 Range  | 
			
				 Resolution  | 
			
				 Accuracy of Readings (45-1KHz)  | 
		
| 
				 5V – 250 V  | 
			
				 0.1 V  | 
			
				 ±1.5% ± 5dgts  | 
		
| 
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 250 V – 600 V  | 
			
				 0.1 V  | 
			
				 ±1.5% ± 5dgts  | 
		
	
	AC Current
(Auto Range, True RMS, Crest Factor<4, CT=1, Max. peak values 880mADC for mA range, 73.5ADC for A range, Overload Protection AC 600A)
| 
				 Range  | 
			
				 Resolution  | 
			
				 Accuracy of Readings (50/60Hz)  | 
		
| 
				 0.5 – 60mA  | 
			
				 0.01mA  | 
			
				 ±0.5% ± 5dgts  | 
		
| 
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 60 – 100mA  | 
			
				 0.01mA  | 
			
				 ±0.5% ± 50dgts  | 
		
| 
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 100 – 600 mA  | 
			
				 0.1 mA  | 
			
				 ±0.5% ± 5dgts  | 
		
| 
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 0.05 – 3 A  | 
			
				 0.001 A  | 
			
				 ±0.5% ± 5dgts  | 
		
| 
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 3 – 30 A  | 
			
				 0.01 A  | 
			
				 ±0.5% ± 5dgts  | 
		
| 
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 30 – 50 A  | 
			
				 0.01 A  | 
			
				 ±1.0% ± 5dgts  | 
		
| 
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 Range  | 
			
				 Resolution  | 
			
				 Accuracy of Readings (45-1KHz)  | 
		
| 
				 0.5 – 60mA  | 
			
				 0.01mA  | 
			
				 ±1.5% ± 5dgts  | 
		
| 
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 60 – 100mA  | 
			
				 0.01mA  | 
			
				 ±1.5% ± 50dgts  | 
		
| 
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 100 – 600 mA  | 
			
				 0.1 mA  | 
			
				 ±1.5% ± 5dgts  | 
		
| 
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 0.05 – 3 A  | 
			
				 0.001 A  | 
			
				 ±1.5% ± 5dgts  | 
		
| 
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 3 – 30 A  | 
			
				 0.01 A  | 
			
				 ±1.5% ± 5dgts  | 
		
| 
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 30 – 50 A  | 
			
				 0.01 A  | 
			
				 ±1.5% ± 5dgts  | 
		
Harmonics of AC Voltage in Percentage
(1 to 99thorder, minimum voltage at 50 or 60 Hz is greater than AC 80V. If the voltage is 0 at 50 or 60 Hz, all the percentage (%) display is 0.)
| 
				 Range  | 
			
				 Resolution  | 
			
				 Accuracy  | 
		
| 
				 1 – 10th  | 
			
				 0.1 %  | 
			
				 ±1% of reading ±1%  | 
		
| 
				 11 – 20th  | 
			
				 0.1 %  | 
			
				 ±5% of reading ±1%  | 
		
| 
				 21 – 50th  | 
			
				 0.1 %  | 
			
				 ±15% of reading ±1%  | 
		
| 
				 51 - 99th  | 
			
				 0.1 %  | 
			
				 ±35% of reading ±1%  | 
		
Harmonics of AC Voltage in Magnitude
(1 to 99thorder, minimum voltage at 50 or 60 Hz is greater than AC 80V)
| 
				 Range  | 
			
				 Resolution  | 
			
				 Accuracy  | 
		
| 
				 1 – 10th  | 
			
				 0.1 V  | 
			
				 ±1% of reading ±7dgts  | 
		
| 
				 11 – 20th  | 
			
				 0.1 V  | 
			
				 ±5% of reading ±7dgts  | 
		
| 
				 21 – 50th  | 
			
				 0.1 V  | 
			
				 ±15% of reading ±7dgts  | 
		
| 
				 51 - 99th  | 
			
				 0.1 V  | 
			
				 ±35% of reading ±7dgts  | 
		
Harmonics of AC Current in Percentage
(1 to 99thorder, min. current at 50 or 60 Hz is greater than 100mA for mA range, and greater than 1A for A range. If the current is 0 at 50 or 60 Hz, all the percentage (%) display is 0)
| 
				 Range  | 
			
				 Resolution  | 
			
				 Accuracy  | 
		
| 
				 1 – 10th  | 
			
				 0.1 %  | 
			
				 ±1% of reading ±1%  | 
		
| 
				 11 – 20th  | 
			
				 0.1 %  | 
			
				 ±5% of reading ±1%  | 
		
| 
				 21 – 50th  | 
			
				 0.1 %  | 
			
				 ±15% of reading ±1%  | 
		
| 
				 51 - 99th  | 
			
				 0.1 %  | 
			
				 ±35% of reading ±1%  | 
		
| 
				 Harmonics of AC Current in Magnitude  | 
		
(1 to 99thorder, min. current at 50 or 60 Hz is
greater than 100mA for mA range, and greater than 1A for A range)
| 
				 Range  | 
			
				 
  | 
			
				 Resolution  | 
			
				 
  | 
			
				 Accuracy  | 
		
| 
				 1 – 10th  | 
			
				 
  | 
			
				 0.01mA /  | 
			
				 reading in mA: ±1% of reading ±2mA  | 
		|
| 
				 
  | 
			
				 
  | 
			
				 0.1mA /  | 
			
				 reading in A: ±1% of reading ±0.3A  | 
		|
| 
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 11 – 20th  | 
			
				 
  | 
			
				 0.001A /  | 
			
				 reading in mA: ±7% of reading ±2mA  | 
		|
| 
				 
  | 
			
				 
  | 
			
				 0.01A  | 
			
				 reading in A: ±7% of reading ±0.3A  | 
		|
| 
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 21 – 50th  | 
			
				 
  | 
			
				 
  | 
			
				 reading in mA: ±15% of reading ±3mA  | 
		|
| 
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 reading in A: ±15% of reading ±0.3A  | 
		|
| 
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 51 - 99th  | 
			
				 
  | 
			
				 
  | 
			
				 reading in mA: ±35% of reading ±3mA  | 
		|
| 
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 reading in A: ±35% of reading ±0.3A  | 
		|
| 
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 Power Factor (PF, ACV > 4V, AC mA > 1mA, AC A > 0.04A, Watt > 50 digits)  | 
		||||
| 
				 Range  | 
			
				 
  | 
			
				 Resolution  | 
			
				 
  | 
			
				 Accuracy  | 
		
| 
				 0.000 – 1.000  | 
			
				 
  | 
			
				 0.001  | 
			
				 
  | 
			
				 ±0.04  | 
		
| 
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 Phase Angle (Φ, 50 or 60Hz)  | 
			
				 
  | 
			
				 
  | 
		||
| 
				 Range  | 
			
				 
  | 
			
				 Resolution  | 
			
				 
  | 
			
				 Accuracy  | 
		
| 
				 -180° to 180°  | 
			
				 
  | 
			
				 0.1°  | 
			
				 
  | 
			
				 ± 2°  | 
		
| 
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 0° to 360°  | 
			
				 
  | 
			
				 0.1°  | 
			
				 
  | 
			
				 ± 2°  | 
		
(THD-F with respect to the first (fundamental) harmonics, min.
value in the range from 45 to 65Hz is greater than AC 80V, greater than 1A for A range, and greater than 100mA for mA range. Calculation is done over 1 to 50thHarmonics. If the voltage or current is 0 in the range from 45 to 65Hz, all the percentage (%) display is 0)
| 
				 Range (45~65Hz)  | 
			
				 Resolution  | 
			
				 Accuracy  | 
		||
| 
				 0.0  | 
			
				 –  | 
			
				 10.0 %  | 
			
				 0.1%  | 
			
				 ± 2%  | 
		
| 
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
		|
| 
				 10.0  | 
			
				 – 40%  | 
			
				 0.1%  | 
			
				 ± 5% of reading ± 5%  | 
		|
| 
				 
  | 
			
				 
  | 
			
				 
  | 
		||
| 
				 40 – 100%  | 
			
				 0.1%  | 
			
				 ± 10% of reading ± 10%  | 
		||
| 
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 100  | 
			
				 –  | 
			
				 999.9%  | 
			
				 0.1%  | 
			
				 ± 20% of reading  | 
		
Peak Value of AC Periodic Voltage (peak value> 10V)or AC Periodic Current (peak
value>10mA for mA range and peak value>0.5A for A range) for continuous
waveform except for square wave:
| 
				 Range  | 
			
				 Sampling Time  | 
			
				 Accuracy of Reading  | 
		
| 
				 50 Hz  | 
			
				 39 μs  | 
			
				 ± 5% ± 30 digits  | 
		
| 
				 
  | 
			
				 
  | 
			
				 
  | 
		
| 
				 60 Hz  | 
			
				 33 μs  | 
			
				 ± 5% ± 30 digits  | 
		
| 
				 
  | 
			
				 
  | 
			
				 
  | 
		
Crest Factor (C.F.) of AC Voltage (peak value > 10V)or AC Current (peak value>10mA for mA range and peak value>0.5A for A range) for continuous waveform except for square wave:
| 
				 Range (45~65Hz)  | 
			
				 Resolution  | 
			
				 Accuracy of Readings  | 
		
| 
				 1.00 – 99.99  | 
			
				 0.01  | 
			
				 ± 5% ± 30 digits  | 
		
	Công ty Cổ Phần Thương mai Dịch vụ và Xuất Nhập Khẩu Bình Minh
	Địa chỉ: Số 9,Ngõ 461 Nguyễn Văn Linh, Q. Long Biên, TP.Hà nội
	Tel: 024-38751616  Hotline: 0904499667
	Email: binhminhthuha@gmail.com
	============================
	Chi nhánh Công ty Cổ Phần Thương mai Dịch vụ và Xuất Nhập Khẩu Bình Minh Tại TPHCM
	Địa chỉ: Số 1331/15/144 Đường Lê Đức Thọ,Phường 14, Q. Gò Vấp, TP.Hồ Chí Minh
	Tel: 024-38751616 và 024-36790555 Hotline: 0988764055
	Email: binhminhthuha@gmail.com
	 
| 
				 AMPE KÌM  | 
			
				 ĐỒNG HỒ ĐO ĐIỆN Đồng hồ đo điện trở cách Kyoritsu Đồng hồ đo điện trở cách Hioki Đồng hồ đo điện trở đất Kyoritsu 
  | 
			
				 MÁY THỬ ĐIỆN ÁP  | 
			
				 MÁY ĐO NHIÊT ĐỘ 
  | 
			
				 MÁY ĐO TỐC ĐỘ  | 
		